Skip to main content

Maintenance

How oftenWhat to do
Every 3 months, and whenever you replace the probeCalibrate the pH sensor. The app reports the probe's health afterwards; replace a probe it reports as worn. See pH.
When EC readings look offClean the EC sensor's metal tips and calibrate it. See EC.
After changing a pump's tubingRecalibrate that pump. See Calibrating a pump.
RegularlyInspect pumps, tubing and cables for wear, kinks and leaks. Check that hose ends still sit where they should, so nothing can siphon.
Before a supply bottle runs dryTop it up. An empty bottle is the most common reason for an automation to stop with Needs attention.

Cleaning​

  • pH probe: rinse in probe storage or cleaning solution, or in pH 4 buffer. Never wipe the glass electrode. See pH.
  • EC and water temperature sensors: mild detergent, or an isopropyl alcohol based disinfectant. Stubborn residue on the EC tips comes off with a scouring pad.
  • Air temperature and humidity sensor: do not clean it.
  • Control unit: keep it dry.

Software updates​

The device updates its own software over Wi-Fi and restarts afterwards. Nothing is needed from you. The app is updated through the app store, and tells you if your version is too old to keep working.